Record
Report numberAN-2026-00018
Issued2026-08-11
StatusValid
SampleBacteriostatic water, 0.9 % benzyl alcohol, vial labelled 10 ml
Batch numberATBAC-10-2607-01
MatrixBacteriostatic water for injection
Determinations performed20
Issued byANALYTON
Certificate fileAN-2026-00018.pdf
01Results 20
As reported on certificate AN-2026-00018
Determination Method Result Specification
Appearance Visual, USP <790>, black and white background, 2000 lx, 5 s Clear, colourless solution free of visible particles; faint aromatic odour of benzyl alcohol Clear, colourless, free of visible particles Pass
Identity of benzyl alcohol GC-MS (EI, 70 eV) and LC-HRMS (ESI+) Retention time 12.90 min matches the standard within 0.05 min; EI spectrum m/z 108 (M+•), 107, 79, 77, NIST library match 96 %; LC-HRMS [M+H]+ 109.0648, deviation 0.0 ppm, [M+Na]+ 131.047 Theoretical monoisotopic 108.0575 Da from C7H8O; [M+H]+ 109.0648 Pass
Benzyl alcohol content GC-FID, DB-624, internal standard 2-phenylethanol; five-point calibration 0.5 to 1.5 % w/v, r² 0.9998 0.91 ± 0.02 % w/v (9.1 mg/ml), 101 % of the labelled 0.9 % (k = 2) 90 to 110 % of label (0.81 to 0.99 % w/v) Pass
Related substances of benzyl alcohol GC-FID; LOD 0.003 %, LOQ 0.01 %, reporting threshold 0.01 % relative to benzyl alcohol Benzaldehyde 0.030 % (RRT 0.85, ± 0.004 %), i.e. 2.7 µg/ml; dibenzyl ether 0.020 % (RRT 1.62); cyclohexylmethanol < 0.01 % (RRT 0.93); total 0.05 % Ph. Eur. 0256, parenteral grade: benzaldehyde <= 0.05 %, cyclohexylmethanol <= 0.10 %, any other <= 0.10 %, total <= 0.3 % Pass
Deliverable volume Gravimetric, USP <697>, density 0.9975 g/ml at 20 °C 10.4 ml (five vials 10.3 to 10.5 ml) Not less than 10.0 ml from each vial Pass
Benzyl alcohol per vial Calculated from content and deliverable volume 94.6 mg per vial (9.1 mg/ml × 10.4 ml) against 90 mg for the labelled 0.9 % in 10 ml, i.e. 105 %; each withdrawn millilitre carries 9.1 mg regardless of overfill 90 to 110 % Pass
pH Potentiometric, USP <791>, 25 °C, KCl bridge for low ionic strength 5.6 ± 0.1 4.5 to 7.0 (USP Bacteriostatic Water for Injection) Pass
Osmolality Freezing point depression, USP <785> 85 ± 3 mOsm/kg (calculated 83 for 9.0 g/l benzyl alcohol); the solution is hypotonic and is not an isotonic vehicle Report only
Conductivity USP <645>, 25 °C 1.1 ± 0.1 µS/cm <= 1.3 µS/cm (stage 1) Pass
Total organic carbon USP <643> Not applicable: benzyl alcohol itself contributes 7.4 mg of carbon per ml, so the test would measure the preservative and not contamination Report only
Density and balance Oscillating U-tube, USP <841>, 20 °C 0.9975 ± 0.0005 g/ml; 9.1 mg/ml benzyl alcohol + 988.4 mg/ml water = 997.5 mg/ml, consistent with the density; no other solute (conductivity 1.1 µS/cm) Report only
Inorganic impurities of the water Ion chromatography; oxidisable substances by permanganate Nitrate < 0.2 ppm, chloride < 0.5 ppm, sulfate < 0.5 ppm, ammonium < 0.1 ppm; oxidisable substances: the pink colour persists Ph. Eur. Water for injections: nitrate <= 0.2 ppm, ammonium <= 0.2 ppm Pass
Bacterial endotoxins Kinetic chromogenic LAL, USP <85>, undiluted sample < 0.005 EU/ml (LOQ 0.005 EU/ml); 0.9 % benzyl alcohol does not interfere <= 0.25 EU/ml (USP Bacteriostatic Water for Injection) Pass
Sterility Membrane filtration with 3 × 100 ml rinse to neutralise benzyl alcohol, USP <71>; 20 vials; fluid thioglycollate 30 to 35 °C and soya bean casein digest 20 to 25 °C No growth after 14 days No growth Pass
Antimicrobial effectiveness USP <51>, category 1 (parenteral), inoculum 10^5 to 10^6 CFU/ml S. aureus 3.4 log reduction at 7 d and > 4.0 at 14 d; P. aeruginosa 3.6 and > 4.0; E. coli 3.8 and > 4.0; C. albicans 1.2 log at 14 d; A. brasiliensis 0.6 log at 14 d; no increase to 28 d. Benzyl alcohol is weak against moulds, which is why the number is reported Bacteria >= 1.0 log at 7 d, >= 3.0 log at 14 d, no increase to 28 d; yeast and mould no increase from the initial count Pass
Residual solvents and volatile impurities Headspace GC, USP <467> Benzene < 0.5 ppm (LOQ 0.5 ppm; precursor of benzyl alcohol); toluene < 1 ppm; methanol, ethanol and acetone each < 10 ppm ICH Q3C: benzene <= 2 ppm (Class 1), toluene <= 890 ppm, methanol <= 3000, ethanol and acetone <= 5000 ppm Pass
Elemental impurities ICP-MS, direct introduction of the solution, USP <233> Cd, Pb < 0.05 µg/l; As, Hg, Co, V < 0.10 µg/l; Ni 0.30 µg/l; for information Al 2.0 µg/l (type I glass) and Fe 1.5 µg/l ICH Q3D(R2) Class 1 and 2A, parenteral PDE per 10 ml daily: Cd 200, Pb 500, As 1500, Hg 300, Co 500, V 1000, Ni 2000 µg/l Pass
Sub-visible particles Light obscuration, USP <788>, mean of 5 vials >= 10 µm 140 per vial; >= 25 µm 6 per vial <= 6000 and <= 600 per container Pass
Container closure integrity Vacuum decay, USP <1207> 0 of 10 vials leaking; bromobutyl stopper, aluminium seal, flip-off cap 0 leaking Pass
Compound screening GC-MS full scan m/z 35 to 400 and LC-HRMS m/z 100 to 1000, both polarities, against reference standards Other preservatives (methylparaben, propylparaben, phenol, m-cresol, chlorobutanol), ethanol, glycerol, propylene glycol and phthalates (DEHP from the closure): none above 1 µg/ml; no sodium chloride (conductivity), so this is not saline Nothing above 1 µg/ml outside the declared composition Pass
02The sample and the run
Controls carried in the same sequence
BlankWater for injection with internal standard injected before the sequence; within RRT 0.80 to 1.30 no peak above 0.02 % of the benzyl alcohol height (maximum 0.004 %) apart from the internal standard
Reference standardUSP Benzyl Alcohol RS (primary) injected at the start and end of the sequence; area ratio difference 0.3 %, retention time within 0.05 min; a commercial USP bacteriostatic water run as control sample within 2 % of its previous value
03Check your copy of the file

This is the SHA-256 of the certificate file exactly as it was issued. Run the command against the file you were sent: if the two strings match, your copy is byte for byte the one this laboratory issued.

A copy that has been re-saved, printed to a new file or passed through a converter will not match, because each of those produces a different file. Compare the certificate exactly as it was sent to you, or the copy downloadable from this page.

ca260d83b76b8d11c4dd032feb58231c2e7e325b9cd62052812ad821e1d9fee4

shasum -a 256 AN-2026-00018.pdf

On Windows: certutil -hashfile AN-2026-00018.pdf SHA256

04Show this result elsewhere

This badge links to this record, so anyone who clicks it lands on the laboratory's own copy of the result.

The same record is available as JSON at /verify/AN-2026-00018-6S6QH3RABS/record.json, so a shop can pull the results into a product page without anyone retyping a number.

Tested by ANALYTON, report AN-2026-00018

<a href="https://analyton.com/verify/AN-2026-00018-6S6QH3RABS/"><img src="https://analyton.com/verify/AN-2026-00018-6S6QH3RABS/badge.svg" alt="Tested by ANALYTON, report AN-2026-00018" width="260" height="70"></a>

Verified against the ANALYTON register. Questions to lab@analyton.com